Check the background count rate of a gamma camera system, under conditions of clinical imaging by a particular radionuclide

-
Mount the collimator to be used and turn the detector head to a scanning position downward.
-
Position the detector head over the center of the table bed.
-
Adjust all the routine settings on the gamma camera console, for a routine acquisition by the radionuclide concerned.
-
Perform a count measurement for a 100sec recording time without any radiation source at the vicinity of the camera.
-
Record the counts and calculate the background count rate.

-
Compare the calculated value of the background count rate with the reference one and background recent values to identify any changes or trends.
-
A rapid increase of the background count rate may be due to a contamination of the camera system -particularly on the collimator face-, the floor, and the bed or to the presence of radioactive sources near the camera.
-
If none of the above is present, the background may be due to electronic "noise".

-
An increase of the background count rate above the 20% of the reference value should lead to further investigation.
-
Record whether the measurements confirm acceptable performance; if not, corrective action should be taken.
|